summary
Comparison of wrist, chest, and eSense ear wearables uses custom wrist and chest devices plus a commercial earable with an MPU-6050, sampled at 50 Hz and transmitted via BLE; six estimation algorithms run against a mannequin gold standard; 20 participants produced over 5,000 compressions; depth MADs are chest 3.4 mm, wrist 4.5 mm, ear 5.9 mm (sensor fusion 3.2 mm) versus a 9.5 mm baseline; rate MAD <1.6 cpm (target 100–120 cpm); limitations: mannequin, flat hard ground, and inability to detect incomplete chest recoil.